OPERATING PARAMETERS OF ATOMIC ABSORPTION SPECTROSCOPY FOR QUANTITATIVE DETERMINATION OF CHROMIUM IN CROME ORES, CROME CONCENTRATES AND TAILINGS
DOI:
https://doi.org/10.46991/PYSU:B/2014.48.1.010Keywords:
Crome ore, microwave, standard solution, acetylene gas mixtureAbstract
The present paper endeavors to define operating parameters of the atomic absorption method of spectroscopy for quantitative determination of Chromium in crome ores, crome concentrates and tailings. A new method of preparing standard solutions (in combination with microwave acid digestion of standard samples) is proposed for designing calibration curves and explicating conditions of supplying acetylene gas mixture and nitrous oxide, which enhance the accuracy of results obtained from the quantitative determination of chromium.
Downloads
Published
2014-03-17
How to Cite
Manukyan, A.V., and R.S. Harutyunyan. 2014. “OPERATING PARAMETERS OF ATOMIC ABSORPTION SPECTROSCOPY FOR QUANTITATIVE DETERMINATION OF CHROMIUM IN CROME ORES, CROME CONCENTRATES AND TAILINGS”. Proceedings of the YSU B: Chemical and Biological Sciences 48 (1 (233):10-14. https://doi.org/10.46991/PYSU:B/2014.48.1.010.